.0 Safety advice

The Aquamatic
II unit is produced according to IEC
TM
601/EN 60601 and listed in the following classes:
• VDE Class of protection 1
• Class IIa (EEC 93/42).
The equipment must only be connected to a power outlet
with a correctly installed ground/earthing contact according
to local and state requlations.
The unit must be installed by a specialist authorized
by GN Otometrics (see chapter 3.3).
The Aquamatic
II may be used in supervised
TM
operation by qualified personnel which have been au
thorised by GN Otometrics and which have been trained
in the operation of
the equipment (IEC 601-1/EN 60601-1).
The mains voltage specified on the type plate must
match the power supply system.
Prior to every use of the equipment, verify that the device
is technically safe and in proper working condition.
Damaged leads must be replaced immediately!
Correct configuration in assembly of country-specific con-
nections
green/yellow:
protective conductor (PE)
blue:
neutral conductor (N)
black resp. brown:
phase (L)
The control panel must be visible and in reach of the
operator. Ensure the equipment is stable prior to use.
In the thermal process for reducing bacterial count hot
water is being ducted through the irrigation handle. Please
do not take the handle from its holder or spray water!
Metal parts can be hot!
Prior to spraying, the water temperature must be checked
by the user!
Switch off main power switch after use and close the
water valve of water supply.
The Aquamatic
II should only be operated in rooms
TM
used for medical purposes, and never in areas subject to
explosion hazards.
1
UK
 All additional equipment, which is connected to the
analog and digital interfaces of the unit, must meet the
requirements of relevant EN specifications (for inst. EN
60950 for data processing equipment and EN 60601 for
electrical medical appliances). In addition, configurations
must satisfy system specification EN 60601-1-1. When
additional equipment is connected to the signal input or
signal output section on the unit, the person making the
connection is deamed "a system configuration operator"
and as such is responsible for meeting the requirements
of system specification EN 60601-1-1. For answers to
additional questions, please contact your local specialist
supplier or GN Otometrics technical service.
 The irrigation jet must not get into contact with contaminated
material.
Only use the warm water jet when a tube tip is installed!
Take caution to avoid injury to the eardrum when intro-
ducing the water jet !
For hygienic reasons jet connections must be changed
after each patient. This also prevents retrograde ger-
mination of the warm water system.
Use only for irrigation of auditory canal !
GN Otometrics is not liable for personal injury and damage to
property if
• original GN Otometrics parts are not used,
• the operating instructions are not completely followed,
• assembly, new settings, alterations, extensions and repairs
have been carried out by personnel not authorised by
GN Otometrics.