7 Conditioning
24
7.2
Automated cleaning and disinfecting
BA 695 / BA 695FBA / BA 888FM can also be cleaned and disinfected
Product name
in suitable cleaning and disinfection equipment.
The cleaning and disinfection equipment used must be approved by its
Cleaning and disinfection equipment
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
✔ BA 695 / BA 695FBA / BA 888FM is conditioned with a cleaning and
Check _ product name
disinfection device.
1. Check whether BA 695 / BA 695FBA / BA 888FM is clean after
conditioning under good lighting (min. 500 lux) and color rendering
index (min. 80 Ra).
2. If they are still dirty, repeat the process.