6556083.book Seite 30 Dienstag, 22. November 2016 11:30 11
7 Conditioning
30
Process
✔ The black nozzle insert is in the spray can tube.
Care _ Fixed connection
1. Take the highspeed handpiece off of the supply hose.
2. Insert the nozzle of the spray can in the driving air tube. Hold the
highspeed handpiece securely with a cloth when doing this.
3. Spray into the turbine for 1-2 seconds.
IMPORTANT: Hold the spray can upright.
4. Wipe any spray that comes out with a disinfection cloth.
5. Repeat the process until the spray is clear.
Operating Instructions BA ULTIMATE
65 56 083 D3560.201.10.03.02