CHANGE ANALYZER FILTER

NOTE: Wear protective gloves and glasses
The gas analyzer filter must be replaced when the machine visualize the "change analyzer
filter" alarm message (usually every 150 analysis):
Press ENTER to proceed with filter substitution, the following screen will be displayed:
Remove the transparent protection by unscrewing the two screws. Press ENTER to continue,
the following screen will be displayed:
Remove the filter. Press ENTER to continue, the following screen will be displayed:
Insert the new filter, make sure that the arrow on the filter is aligned with the one on the
analyzer. Press ENTER to continue, the following screen will be displayed:
INSTRUCTIONS MANUAL
CHANGE ANALYZER FILTER
It is necessary
to replace analyzer filter.
Press ENTER to proceed
with filter substitution.
Press ESC to exit.
CHANGE ANALYZER FILTER
Remove the transparent protection
by unscrewing the two screws.
Press ENTER to continue
CHANGE ANALYZER FILTER
Remove the filter.
Press ENTER to continue
CHANGE ANALYZER FILTER
Insert the new filter, make sure that
the arrow on the filter is aligned
with the one on the analyzer.
Press ENTER to continue
72
DD/MM/YY XX:XX PM
ESC
ENTER
DD/MM/YY XX:XX PM
ESC
ENTER
DD/MM/YY XX:XX PM
ESC
ENTER
DD/MM/YY XX:XX PM
ESC
ENTER