Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 48
Directions for Use

Troubleshooting

Condition
Power does not come on. Power Cord
No image on monitor.
Lamp will not light.
Printer does not
respond properly to
handle buttons.
Check
Attachment of unit to handle
Attachment of unit to handle and door
to unit
Fuses
Wall outlet
Monitor/other peripheral devices
Cable connections
Monitor input selection
Lamp housing
Lamp
Lamp access door
Power cord and printer
Printer
Maintenance
Action
Check connections at power supply and
wall outlet.
Check colposcope unit's alignment with
handle.
Make sure lamp access door and handle
are properly attached.
Remove fuse panel & replace blown fuse
with T1.00L 250V 1A time lag (Welch Allyn
part number # 488307-9).
Plug the power cord into a wall outlet
known to work.
Make sure power is on for all devices.
Make sure all video cables are connected
properly.
Make sure correct input selection is made
on the front panel.
Make sure lamp assembly is installed
properly.
Replace lamp.
Make sure lamp access door is properly
closed.
Unplug power cord from wall outlet, then
plug cord back in to outlet. At the same
time, turn printer off and then on again.
Make sure printer is on and is properly
connected.
31

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

88001a89000a8800789001a88002a88004a ... Afficher tout

Table des Matières