Directions for use
2
Directions for use
2.1 Intended use
The intended use is 24-hour blood pressure measurement and pulse wave analysis (PWA).
®
The Mobil-O-Graph
is only to be used under medical supervision by medical personnel.
2.2 Improper use
▪
The Mobil-O-Graph
pressure described herein.
▪
Due to the risk of strangulation by the tube and cuff, the Mobil-O-Graph
unsound mind and must not be left within reach of unsupervised children.
▪
The Mobil-O-Graph
▪
The Mobil-O-Graph
in intensive care units.
▪
The Mobil-O-Graph
Note
®
The Mobil-O-Graph
has not been tested for use on pregnant women or those suffering from preeclampsia.
2.3 Essential performance features
The essential performance features are defined as blood pressure measurement with:
▪
Error tolerances of the pressure gauge and measurement results within the required limits according to
IEC 80601-2-30
▪
Maximum change value in blood pressure determination according to IEC 80601-2-30
▪
Power delivery (pressure supply to the cuff) within the set limits according to IEC 80601-2-30, EN 1060-3
▪
An error is issued if successful blood pressure measurement is not possible.
The device does not trigger any alarms in the sense of IEC 60601-1-8 and is not intended for use together with
HF surgical devices or for clinically monitoring patients, e.g. in an intensive care unit.
70
®
must not be used for any purpose other than the process of measuring blood
®
is not intended for use on babies and children under the age of 3.
®
must not be used for alarm-triggering blood pressure monitoring during operations or
®
must not be used on aeroplanes!
®
must not be used on patients of