Guidelines and manufacturer's declaration – electromagnetic immunity
®
The Mobil-O-Graph
is intended for use in the electromagnetic environment specified below. The
customer or Mobil-O-Graph
Measurement of
interference emissions
Radiated RF
disturbance according
to
IEC 61000-4-3
126
®
user should ensure that it used only in such an environment.
IEC 60601-1 test level
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1 kHz sine)
PM; 18 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz
Compliance level
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1 kHz sine)
PM; 18 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz