Guidelines and manufacturer's declaration – electromagnetic immunity
The blood pressure measuring device Mobil-O-Graph
below. The customer or user of the blood pressure measuring device Mobil-O-Graph
used in such an environment.
Immunity test
IEC 60601-1 test level
Electrostatic
+- 8 kV contact discharge
discharge (ESD)
+- 15 kV air discharge
according to IEC
61000-4-2
Fast transient
± 1 kV
electrical
100 kHz repetition rate
disturbance/bursts
according to IEC
61000-4-4
Surge voltages
(surges) according to
IEC 61000-4-5
®
is intended for use in the environment specified
Compliance level
+- 8 kV contact discharge
+- 15 kV air discharge
± 1 kV
100 kHz repetition rate
Not applicable
Appendix
®
should ensure that it is
Electromagnetic environment
– guidelines
Floors should consist of wood
or cement or be covered with
ceramic tiles. If the floor
consists of synthetic materials,
relative humidity must be at
least 30%.
®