Troubleshooting
Error message
Err 7
Err 8
Err 9 +
possible
continuous alarm
until a button is
pressed
Err 10 +
Continuous
alarm until a
button is pressed
106
Possible cause
The memory of the blood pressure
measuring device is full (a
maximum of 300 measurements
and events can be stored or in the
case of PWA a maximum of 260)
Measurement cancelled by
pressing a button
1. Residual pressure inside the
cuff
2. Zero point comparison could
not be performed successfully.
1. Serious error due to pressure
build-up outside the measuring
process (pump has been switched
on against orders)
2. These error messages all show
a severe error in the program
code.
Measures
Delete the data in the blood
pressure monitor but make sure
that the data has been saved (see
chapter 3.4.2 "Clearing the
memory").
Repeat the measurement
1. Wait for the cuff to deflate
completely.
2. Send the device immediately to
your specialist supplier or directly
to IEM GmbH for checking.
Send the device immediately to
your specialist supplier or directly
to IEM GmbH for checking and
repairs.