Error message
The evaluation unit is not
reacting to data transfer but
co is shown on the display.
®
The Mobil-O-Graph
takes
measurements every two
minutes.
The desired log cannot be
set with the button
combination.
The blood pressure monitor
cannot be switched on.
Possible cause
1. Data transmission cable has not
been inserted properly into the
PC.
2. Also see Err 4
Log 9 is set in the Mobil-O-
®
Graph
.
There are still readings from the
last patient in the memory
1. The power packs or batteries
have been incorrectly inserted
2. The power pack or battery
voltage is too low
3. The display is faulty
Troubleshooting
Measures
1. Check whether the 9-pin plug of
the data transmission cable is
securely located in the device's
interface socket.
2. Also see Err 4
Set log 1 or 2 (see chapter 3.4.5
"Specifying the desired
measurement log")
Delete the data in the blood
pressure monitor but make sure
that the data has been saved (see
chapter 3.4.2 "Clearing the
memory")
1. Reinsert both power packs or
batteries and ensure correct
polarity.
2. Replace the power packs or
batteries
3. Send the device to your
specialist supplier or directly to IEM
GmbH for repair.