Some errors are combined with a continuous alarm for safety reasons. The continuous alarm can be ended
by pressing any button. If there is residual pressure inside the cuff, open the cuff immediately.
®
5.4 Mobil-O-Graph
error description
Error message
Err 1
Err 2
Possible cause
1. The patient has severe
arrhythmia
2. Arm moved during
measurement
3. Insufficient valid pulse rate
detected
1. Arm moved during
measurement
2. Cuff does not fit the arm snugly
Troubleshooting
Measures
1. Blood pressure monitor not
applicable
2. Keep the arm still during
measurement
3. Place the cuff on the arm again
1. Keep the arm still during
measurement
2. Check the seating of the cuff and
that of the device
EN
103