ATTENTION
Damage to the device
▪
Liquid must not get into the device. If you think that liquid has entered the device during cleaning or
use of the device, the device must not be used any more.
▪
If the device is exposed to moisture, switch the device off and remove the batteries. Be sure to inform
technical support or send the device to your specialist supplier or manufacturer.
▪
The device must not be used in the vicinity of a magnetic resonance imaging apparatus or in the direct
proximity of another electrical medical device.
▪
®
The Mobil-O-Graph
▪
Do not drop the device and do not place objects on top of it.
▪
Do not use the device directly adjacent to other devices or stacked with other devices, as this may
result in malfunction. If it nevertheless becomes necessary to use the device in the manner described
above, this device and the other devices should be monitored to ensure that they are functioning
correctly.
▪
The use of components other than those supplied with the product may lead to measurement errors,
as alternative transformers and cables, for example, may increase electromagnetic interference
emissions or reduce electromagnetic immunity. You should therefore only use IEM accessories.
▪
The device must not be connected to a PC or other device when it is still electrically connected to a
patient.
▪
Measurement can be interrupted at any stage by pressing any button. This deflates the cuff and the
device can be removed.
▪
Remove the batteries or power pack from the battery compartment as soon as they have run out of
charge or if you do not expect to use the measuring device for a long period of time.
▪
The cuff and tube are made of a material that does not conduct electricity. They therefore protect the
device against the effects of discharging a defibrillator. In the event of discharging a defibrillator, the
is not suitable for simultaneous use with high-frequency surgical devices.
Directions for use
EN
77