Warm-up Time
To ensure accurate results are delivered,
0
the balance must warm up for at least
30 minutes after initial connection
to the power supply. Only after this
time will the device have reached the
required operating temperature.
30
When a verified balance of the accuracy class k
for use in legal metrology is connected to the mains
power, it must warm up for at least one hour before
operation.

3.5 Leveling

Leveling the balance compensates for slant or
unevenness at the place of installation. The balance
must be perfectly horizontal to ensure consistent,
reproducible weighing results.
The balance needs to be re-leveled and then adjusted
each time its setup location is changed.
Secura Balances:
When "Level" appears on the display, you must level the balance.
t
t
t
t
t
t
Quintix
Balances:
®
t
t
− Air bubble at "12 o'clock": turn the
− Air bubble at "3 o'clock": turn the
− Air bubble at "6 o'clock": turn the
− Air bubble at "9 o'clock": turn the
6
Installation and Operating Instructions Manual Secura
1) First of all, plug the AC adaptor
into the socket at the back of your
balance.
2) Then connect the AC adaptor to
the wall outlet (main voltage).
Make sure that the weighing pan is
unloaded.
Press "LEVEL".
Follow the instructions on the
interactive display.
Turn the two feet as shown in the
illustration until the air bubble is
centered within the circle of the
level indicator.
two feet clockwise.
left foot clockwise and the right
foot counterclockwise.
two feet counterclockwise.
left foot counterclockwise and the
right foot clockwise.
, Quintix
®
Instructions for Use in Legal Metrology
Additional Display Elements:
Metrology line with additional information about minimum load
(Min) and verification scale interval (e).
Warning symbol to indicate calculated values,
negative gross values or values that are smaller than
the minimum sample quantity (SQmin).
Setup Menu "Settings":
On balances verified for use in legal metrology,
not all functions and settings are available.
Interface Specification (Data Output)
Settings "PC – SBI" and "USB RS232 SBI" for the identification of
digits not verified for use in legal metrology: In the setting "SBI",
there is no automatic identification of display digits not verified for
use in legal metrology. Corresponding measures or settings must be
carried out on the peripheral device.

3.6 Anti-theft Locking Device

®
A "Kensington" key lock can be
t
t
installed at the fastening point on
the back of the balance if required.
The balance can also be secured
t
t
at the installation location with
a chain and lock, for example.