Use the correct mains adapter for your mains power
t
supply:
Mains adapter sets
Bag
Region/country
YEPS01-PS1
− USA and Japan (US+JP)
− Europe (EU)
− United Kingdom (GB)
YEPS01-PS6
− Argentina (AR)
− Brazil (BR)
− Australia (AU)
− South Africa (ZA)
YEPS01-PS7
− China (CN)
− India (IN)
− Korea (KR)
t
t
t
y
Removing/Replacing the Mains
Adapter
t
t
Connecting a PC/Notebook
Insert the USB cable into a USB socket on a PC or laptop.
t
Connecting an AC Adapter (Optional)
Insert the USB cable into the AC adapter.
t
Plug the AC adapter into the wall outlet (supply voltage).
t
Insert the power plug adapter
into the power supply. The
grooved button must be facing
upwards.
Push the power plug adapter as
far as you can until it clicks into
place.
Check whether the power plug
adapter is securely locked in
place by pulling it gently.
If the power plug adapter does
not move, it is locked in place.
Press the grooved button from
above and pull back on the
mains adapter.
Push the power plug adapter
out of the power supply and
remove it.

3.7 Anti-theft Locking Device

t

3.8 Leveling

Leveling for Models Verified for Use in Legal Metrology
Only
Leveling the scale compensates for slant or unevenness at the
place of installation. The scale must be perfectly horizontal to
ensure consistent, reproducible weighing results.
The scale needs to be re-leveled and then adjusted if
necessary each time its setup location is changed.
t
12*
9*
− Air bubble at "12 o'clock:" Turn
6*
− Air bubble at "3 o'clock:" Turn the
* o'clock
− Air bubble at "6 o'clock:" Turn
− Air bubble at "9 o'clock:" Turn the

3.9 Warm-up Time

Using a Scale Verified for Use in Legal Metrology:
Ensure that there is a warm-up time of at least 24
hours after initial connection to the power supply.
To ensure accurate results are
0
delivered, the scale must warm up
for at least 30 minutes after initial
connection to the power supply.
Only then will the device have
reached the required operating
30
temperature.
Installation Instructions EVO1Y | LAB1Y
English
If required, secure the scale at
the back.
Turn the two leveling feet as
shown in the illustration until
the air bubble is centered within
the circle of the level indicator.
both feet clockwise.
left foot clockwise and the right
foot counterclockwise.
both feet counterclockwise.
left foot counterclockwise and
the right foot clockwise.