Bien Air CHIROPRO L Instructions D'utilisation page 22

Table des Matières

Publicité

Default values
Endo* : default values
The table shows the default operating values
for the endodontics sequence.
ENDODONTICS
OPEN PULP CHAMBER
1:5
7.20 mNm
100'000 RPM
ENDO FILE 1
1:1
30.2 mNm
250 RPM
ENDO FILE 2
1:1
10.1 mNm
250 RPM
ENDO FILE 3
1:1
14.9 mNm
250 RPM
ENDO FILE 4
1:1
20.2 mNm
250 RPM
ENDO FILE 5
1:1
30.2 mNm
250 RPM
ENDO FILE 6
1:1
20.2 mNm
250 RPM
ENDO FILE 7
1:1
14.9 mNm
250 RPM
ENDO FILE 8
1:1
14.9 mNm
250 RPM
ENDO FILE 9
1:1
10.1 mNm
250 RPM
130
Surgery* : default values
The table shows the default operating values for 4 types of surgical operations proposed
by the system, namely:
• Root resection
• Extraction of wisdom teeth
• Sinus raising
• Free program, available to the user.
APICAL RESECTION
TOOTH EXTRACTION
1:5
0.72 Ncm
100'000 RPM
SURGERY
SINUS LIFT
1:2
2.40 Ncm
0.72 Ncm
80'000 RPM
50'000 RPM
FREE
1:5
1:5
0.72 Ncm
100'000 RPM
*only CHIROPRO L

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Chiropro

Table des Matières