Follow.the.manufacturer's.instructions.to.sterilize.the.device(s).and.tray..Select.the.
appropriate.cycle.
Wrapping
Temperature
Time
Dry Time
EN-8
STERRAD®
.
STERRAD®
100S and
.
STERRAD®
NX
375-708-500.Formula.180.
Shaver.Handpiece.
375-704-500.Formula..
Shaver.Handpiece
375-701-500.Formula.
Shaver.Handpiece
Autoclave (Steam)
Gravity
double
132°C
(270°F)
20.minutes
60.minutes
Warning:.Drying.time.depends.on.several.
..
variables,.including:.altitude,.humidity,.type.of.
wrap,.preconditioning,.size.of.chamber,.mass.of.
load,.material.of.load,.and.placement.in.chamber..
Users.must.verify.that.drying.time.set.in.their.
autoclave.yields.dry.surgical.equipment.
Warning:.These.instructions.were.developed.
.
using.the.guidance.from.AAMI.TIR.12,.ISO.17665.
and.AAMI.ST79.and.Stryker.recommends.users.
observe.these.standards.
STERRAD® Advanced NX
275-601-500.TPS.Small.Joint.
Shaver
Prevacuum
double
132°C
(270°F)
4.minutes
50.minutes
"Flash"
Prevacuum
132°C
(270°F)
4.minutes