NSK ENDO-MATE DT2 Mode D'emploi page 21

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 91
Operation Procedure
AUTO REVERSE OFF
The motor handpiece stops without reverse rotation when the torque limit value is exceeded� In this
instance, the LCD display shows "- - -" and the rotation speed alternately� To rotate it again (forward
rotation), remove the load and press the ON/OFF button twice�
AUTO STOP
The motor handpiece rotates in reverse when the torque limit value is exceeded� When the load is
removed, the motor handpiece stops� If you want it to rotate (forward rotation) again, re-press the ON/
OFF button�
Load within the torque
limit value
Forward Rotation
● When reverse rotation is set, this Auto Reverse Function is not activated�
● If the battery goes down (when charge indicator mark "
reach the preset torque limit value (in the case of operation by the battery)� In this case, this auto
reverse function will not be activated� When high torque is required, use the AC adaptor or use
this product in a condition where the battery is not significantly consumed (when charge indicator
mark "
,
" appears)�
● If a load is continually applied to the motor handpiece, it may automatically stop to prevent
overheating (the torque limit value indicates "O�H" on the LCD display�) In this case, leave the
motor handpiece for a while until it cools down� When the display returns to normal, the motor
handpiece can be used�
Further loading
beyond the preset
torque limit value
Reverse Rotation
NOTICE
,
20
If the load is kept applied, the file
will rotate in reverse.
When the load is removed,···
Stop
" appears), actual load may not

Publicité

Table des Matières
loading