6
Clean the surface of the contra angle head
under running water with a soft bristled
brush for more than 20 seconds�
Water temperature: ≤38°C
Water quality:
Water amount:
7
Clean the file insertion hole for more than 20
seconds with a wire (ISO 16409 size 4 (wire
thickness 0�7mm)) under running water�
Water temperature: ≤38°C
Water quality:
Water amount:
[Cleaning and Disinfection of Interior]
● Before cleaning and disinfecting the interior, be sure to clean the casing�
Water available as
drinking water
≥3�5 L/min
Water available as
drinking water
≥3�5 L/min
CAUTION
8
Wipe the residual moisture on the surface of
the contra angle head with a dry cloth�
9
After cleaning, under appropriate lighting
(≥500lx), inspect the contra angle head for
contamination�
If any visible contamination remains, repeat
the process from the procedure No� 6 until it
is visually clean�
≥ 500lx