13. Maintenance
Check the system joint for wear and functionality every 6 months. Also check the functionality after every
maintenance carried out.
Joint Component
extension stop face
locking pawl
spring guide
pressure spring in upper
part/lower part
sliding bushing
sliding washer
bearing nut
Secure the screws for the joint's upper part with the torque corresponding to the system width and
LOCTITE® 243 medium strength during every maintenance (see paragraph 8.6). Remove all adhesive residues
first.
13.1 Replacing the Sliding Washers
Sliding washers are available in different thicknesses (e.g. GS2413-040 is 0.40mm thick). Each thickness has
a different marking (fig. 21). You will find the article numbers of the premounted sliding washers on the back
page of these instructions for use.
0.40mm
13.2 Dirt Removal
The NEURO LOCK H
O system knee joint is suited for a usage in wet areas. Nevertheless, dirt must be removed
2
when necessary and during regular maintenance. For this purpose, disassemble the system joint and clean the
soiled system components with a dry cloth.
In order to optimise the lifespan, we recommend rinsing the orthosis with water, especially after using it in
salt water, chlorine water and sand.
Problem
wear
wear
wear
fatigue of spring force
wear
wear
wear
0.45mm
0.50mm
Measure
replacing extension stop
replacing locking pawl
replacing spring guide
replacing pressure spring
replacing sliding bushing
replacing sliding washer, see paragraph 12.1
replacing bearing nut
0.55mm
0.60mm
fig. 21
27