Contra-
The bed is not suitable for use in the following situations:
indications
* Application Environments are defined in
IEC 60601-2-52.
The bed is not suitable for patients under 40kg in weight.
The maximum recommended patient weight is 185kg.
185 kg/
407 lb
The safe working load (SWL) of the bed is 250kg.
250 kg/
550 lb
The safe working load is calculated as follows (in accordance
with IEC 60601-2-52):
Maximum patient weight........................................... 185kg
Mattress .................................................................... 20kg
Accessories (including attached loads) ...................... 45kg
TOTAL
WARNING
If the combined weight of the mattress and accessories exceeds
65kg, the maximum patient weight must be reduced accordingly.
The recommended patient size is; weight equal or above 40kg,
height between 146cm and 190cm and BMI equal or above 17.
At the discretion of the carer, patients taller than 190cm may be
146
cm
BMI 17
accommodated by extending the bed -
40
kg
adjustment" on page 24.
exceed the
• A domestic area, i.e. home healthcare (*Application
Environment 4).
• Outpatient care (*Application Environment 5).
.................................................................. 250kg
"In-bed length" shown on page 50.
refer to "Bed length
Ensure that the patient's height does not
11