Appearance Of The Heart Arrhythmia Indicator For Early Detection; Data Memory; Battery Indicator And Battery Change; Low Battery - Microlife BP 3AG1 Mode D'emploi

Masquer les pouces Voir aussi pour BP 3AG1:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 10
4. Appearance of the Heart Arrhythmia Indicator for
early Detection
This symbol AK indicates that certain pulse irregularities were detected
during the measurement. In this case, the result may deviate from your
normal blood pressure – repeat the measurement. In most cases, this is
no cause for concern. However, if the symbol appears on a regular basis
(e.g. several times a week with measurements taken daily) we advise you
to tell your doctor. Please show your doctor the following explanation:
Information for the doctor on frequent appearance of the
Arrhythmia indicator
This instrument is an oscillometric blood pressure monitor that
also analyses pulse frequency during measurement. The instru-
ment is clinically tested.
The arrhythmia symbol is displayed after the measurement, if pulse
irregularities occur during measurement. If the symbol appears
more frequently (e.g. several times per week on measurements
performed daily) we recommend the patient to seek medical advice.
The instrument does not replace a cardiac examination, but
serves to detect pulse irregularities at an early stage.

5. Data Memory

This instrument always stores the last result at the end of the
measurement. To call up the value, press and hold down the
ON/OFF button 1 (the instrument must be switched off first). All
display elements are now shown. Release the button when you
see the stored reading value and the letter «M».

6. Battery Indicator and Battery change

Low battery

When the batteries are approximately ¾ empty the battery symbol
AM will flash as soon as the instrument is switched on (partly filled
battery displayed). Although the instrument will continue to
measure reliably, you should obtain replacement batteries.
Flat battery – replacement
When the batteries are flat, the battery symbol AM will flash as soon
as the instrument is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment 5 at the back of the instrument.
4
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.

Which batteries and which procedure?

Use 4 new, long-life 1.5V, size AA batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the instrument is not going to be used
for a prolonged period.

Using rechargeable batteries

You can also operate this instrument using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
instrument as they may become damaged (total discharged as
a result of low use of the instrument, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the instrument for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.

7. Using a Mains Adapter

You can operate this instrument using the Microlife mains adapter
(DC 6V, 600mA).
Only use the Microlife mains adapter available as an orig-
inal accessory appropriate for your supply voltage, e.g. the
«Microlife 230V adapter».
Ensure that neither the mains adapter nor the cable are damaged.
1. Plug the adapter cable into the mains adapter socket 4 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is consumed.

Publicité

Table des Matières
loading

Table des Matières