Taking A Blood Pressure Measurement Using This Instrument; How Not To Store A Reading; Appearance Of The Heart Arrhythmia Indicator For Early Detection - Microlife BP A200 Mode D'emploi

Masquer les pouces Voir aussi pour BP A200:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 10
• The individual results are not displayed. Your blood pressure
will only be displayed after all 3 measurements are taken.
• Do not remove the cuff between measurements.
• If one of the individual measurements was questionable, a
fourth one is automatically taken.
3. Taking a Blood Pressure Measurement using this
Instrument
Checklist for taking a reliable measurement
1. Avoid activity, eating or smoking immediately before the
measurement.
2. Sit down for at least 5 minutes before the measurement and
relax.
3. Always measure on the same arm (normally left).
4. Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do
not interfere with the cuff if they are laid flat.
5. Always ensure that the cuff is positioned correctly, as shown
in the pictures illustrated on the short instruction card.
• Fit the cuff closely, but not too tight.
• Make sure that the cuff is 2 cm (0.75 inch) above your
elbow with the tube on the inside of your arm.
• Support your arm so it is relaxed.
• Ensure that the cuff is at the same height as your heart.
6. Press the ON/OFF button 1 to start the measurement.
7. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
8. When the correct pressure is reached, the pumping stops
and the pressure falls gradually. If the required pressure was
not reached, the instrument will automatically pump some
more air into the cuff.
9. During the measurement, the heart symbol BT flashes in the
display and a beep sounds every time a heartbeat is detected.
10. The result, comprising the systolic AM and the diastolic AN
blood pressure and the pulse AO is displayed and longer beep
is heard. Note also the explanations on further displays in
this booklet.
4
11. When the measurement has finished, remove the cuff.
12. Enter the result in the enclosed blood pressure pass and
switch off the instrument. (The monitor does switch off auto-
matically after approx. 1 min.).

How not to store a reading

Press the ON/OFF button 1 while the reading is being displayed.
Keep the button pressed until «M» AS is flashing and then release
it. Confirm by pressing the M-button.
You can stop the measurement at any time by pressing the
ON/OFF button (e.g. if you feel uneasy or an unpleasant
pressure sensation).
If the systolic blood pressure is known to be very high,
it can be an advantage to set the pressure individually. Press
the ON/OFF button after the monitor has been pumped up to
a level of approx. 30 mmHg (shown on the display). Keep the
button pressed until the pressure is about 40 mmHg above
the expected systolic value – then release the button.
4. Appearance of the Heart Arrhythmia Indicator for
early Detection
This symbol AP indicates that certain pulse irregularities were
detected during the measurement. In this case, the result may
deviate from your normal blood pressure – repeat the measure-
ment. In most cases, this is no cause for concern. However, if the
symbol appears on a regular basis (e.g. several times a week with
measurements taken daily) we advise you to tell your doctor.
Please show your doctor the following explanation:
Information for the doctor on frequent appearance of the
Arrhythmia indicator
This instrument is an oscillometric blood pressure monitor that
also analyses pulse frequency during measurement. The instru-
ment is clinically tested.
The arrhythmia symbol is displayed after the measurement, if
pulse irregularities occur during measurement. If the symbol
appears more frequently (e.g. several times per week on
measurements performed daily) we recommend the patient to
seek medical advice.
The instrument does not replace a cardiac examination, but
serves to detect pulse irregularities at an early stage.

Publicité

Table des Matières
loading

Table des Matières