Calibration Data Memory (Read Cal); Alarm ("Al."); Real-Time Clock ("Cloc") - Lovibond SD 310 Oxi Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 58
13.2 Calibration data storage (rEAd CAL)
The last 16 calibrations with date and results are stored in the device and can be called up.
Display the calibration data memory:
Saved calibration data can be read in the GSOFT3050 software and can be observed in the
device display itself:
Press and hold for two seconds:
The following appears in the display:
Press until the following appears:
Press briefly: Switch between
or

14. Alarm ("AL.")

There are three possible settings:
Off (AL.oFF), On with sound (Al.on), On without sound (AL.no.So).
In the following cases an alarm is issued when the alarm function is active (on or no.So.):
-
Lower alarm threshold (Al. Lo) undercut
-
Upper alarm threshold (Al. Hi) exceeded.
-
Sensor error
-
Low battery (bAt)
-
Err.7: System error (is always signaled with sound)
In the event of an alarm, the "PRIO" flag is marked in the device response with interface
accesses.
15. Real time clock ("CLOC")
The real time clock is required for the temporal assignment of the logger data and the cali-
bration time point. Therefore, check the settings as necessary.
with 1 and 2-point calibration is SL.1 = SL. 2
*1)
With 3-point calibration, independent values are determined for both ranges.
52
- ELE[ = total evaluation in %
- SL. 1 = pitch 0% air
- SL. 2 = pitch air – 100%
- Date+time display of the data set
Switch between the calibration data sets
Close the display of the calibration data sets
or
read cal. = "Read calibration data"
*1)
*1)
(configuration level)
SD_310_2 07/2018

Publicité

Table des Matières
loading

Table des Matières