Sensor Evaluation (Ele - Lovibond SD 310 Oxi Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 58
As long as the display is flashing, no valid reference has been detected.
As soon as the measurement values are constant, the calibration of the second point is
completed automatically.
With 2-point calibration, the calibration is completed and a sensor evaluation is displayed in
10% increments (display briefly shifts to xx% ELE[).
With 3-point calibration, you will be prompted to create the next reference (possible
!
reference starts flashing)
4. Calibration point 3: (Pt.3)
appears in the display as well as the reference you want to create.
As soon as the measurement values are constant, the calibration is completed automatically
and then a sensor evaluation is displayed in 10% increments (display briefly shifts to xx% ELE[).
If error messages appear during calibration, follow the instructions at the end of these
instructions. If the sensor cannot be calibrated after a long period, at least one of
the measurement values (oxygen concentration, temperature) is unstable. Check the
measurement system!

9.3 Sensor evaluation (ELE[)

Examine the sensor evaluation:
The sensor evaluation of the last successful calibration is displayed.
The evaluation is done in 10% increments. 100% is the optimum status. Lower values indicate
that the sensor is close to the end of its service life.
Comment: Too low measurements can be caused by incorrectly measured air pressure.
SD 310_2 07/2018
Briefly press CAL
display briefly changes to xx% ELE[
47

Publicité

Table des Matières
loading

Table des Matières