Lovibond SD 310 Oxi Mode D'emploi page 40

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 58
Menu
Parameter
SET
Adjusts the measurement channels
(OR R
0FFS
°C or °F
S[AL
°C or °F
0FFS
hPa
SET
Set Alarm: Adjustment of the alarm function
AL. 1
AL
AL.,N
A 1.LO
A 1.H,
AL. 2
A2.LO
(*)
If there is data in the logger memory, parameters marked with (*) cannot be opened.
If they should be changed, the data must first be deleted!
40
Values
Meaning
-5.0 °C to 5.0 °C
The zero point of the temperature
or
measurement is offset by the set
-9.0 °F to 9.0 °F
value. The sensor differences and the
measurement device differences can
both be offset
OFF
Zero point offset is deactivated
(=0.0°)
-5.00 - 5.00 % The pitch of the temperature
measurement is modified by this
factor (in %) to offset the sensor and
measurement device differences.
Pitch correction is deactivated (=0.00)
OFF
-20 - 20 hPa
The zero point of the absolute pressure
measurement is offset by the set value.
Sensor differences can thus be offset.
OFF
Zero point offset is deactivated (=0.0°)
ON / NO.SO
Oxygen monitoring: Alarm on with
horn / alarm on without horn
OFF
No alarm monitoring for oxygen
[ONC
Oxygen alarm channel: Concentration
in %
P. 02
Oxygen alarm channel: Partial pressure
in hPa or mmHg
e . g . 0 . 0 t o
Oxygen min. alarm threshold (not with
100.0 %
AL. 1. oFF)
e . g . 0 . 0 t o
Oxygen max. alarm threshold (not with
100.0 %
AlL 1. oFF)
ON / NO.SO
Temperature measurement alarm on
with horn / alarm on without horn
OFF
No alarm function for temperature
measurement
-5.0 to +50.0
Temperature min. alarm threshold (not
°C
with Al. 2. oFF)
see
SD_310_2 07/2018

Publicité

Table des Matières
loading

Table des Matières