5.4 Mechanical neutral or mild-alkaline cleaning and thermal disinfection

Machine type: Single-chamber washer/disinfector without ultrasound
Stage
Step
I
Prerinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfecting
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated (demineralized) water
Check visible surfaces for residues after mechanical cleaning/
disinfecting.
Repeat the cleaning process if necessary.
Note
Ultrasonic treatment can be applied in manual processing as well as in
mechanical processing. Temperatures of up to 120 °C are permissible for
mechanical drying with hot air.
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD–W
>10/50
1
FD–W
90/194
5
FD–W
-
-
-

5.5 Inspection, maintenance and checks

Note
The sterile containers may be tested and repaired only by persons with the
appropriate training, expertise or experience.
The seals in the container lids were tested according to DIN EN 868-8,
Annex G, for a minimum service life of 5 000 sterilization cycles.
In case of visible damage, change the seal immediately.
Have the lid sent to Aesculap, see Technical Service.
If necessary, lightly lubricate moving metal parts, e.g. latch hinges,
with sterilizable and steam-permeable maintenance oil (e.g. STERILIT®
oil spray JG600 or STERILIT® I drip feed oiler JG598).
Chemical
-
Neutral:
BBRAUN HELIMATIC CLEANER
neutral, pH-neutral,
process solution 0.5 %
Mild-alkaline:
- Concentrate:
pH = 9.5
< 5 % anionic tensides
- 0.5 % solution
Alkaline:
Processing possible at pH up to 10.5, provided the cleaning
agent is manufacturer-approved for container cleaning.
Note
Application of mild-alkaline or alkaline cleaning agents
may cause discoloration of colored anodized aluminum,
which however will not affect the functionality of the
product.
Especially for PrimeLine it must be ensured that the
surface is rinsed without leaving any residues.
Other process parameters may be feasible with agreement
by the hospital hygienist.
According to mechanical cleaning program
9