Data Memory; Viewing The Stored Values; Memory Full; Clearing All Values - Microlife BP A50 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 8

5. Data Memory

At the end of each measurement, this device automatically stores
the result.

Viewing the stored values

Press the M-button 3 briefly, when the device is switched off. The
display first shows the last stored result.
Pressing the M-button again displays the previous value. Pressing
the M-button repeatedly enables you to move from one stored
value to another.

Memory full

When the memory has stored 14 results, the memory is full.
From this point onwards, a new measured value is stored
by overwriting the oldest value.

Clearing all values

All memory data are cleared when batteries are taken out
of the device.

6. Battery Indicator and Battery change

Low battery

When the batteries are approximately ¾ empty the battery symbol
AO will flash as soon as the device is switched on (partly filled
battery displayed). Although the device will continue to measure
reliably, you should obtain replacement batteries.
Flat battery – replacement
When the batteries are flat, the battery symbol AO will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment 7 at the back of the device.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.

Which batteries and which procedure?

Use 2 new, long-life 1.5 V, size AAA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.
Using rechargeable batteries
You can also operate this device using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
4
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.

7. Error Messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description Potential cause and remedy
«ERR 1» Signal too
The pulse signals on the cuff are too weak.
weak
Re-position the cuff and repeat the
measurement.*
«ERR 2» Error signal During the measurement, error signals were
detected by the cuff, caused for instance by
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«ERR 3» No pres-
An adequate pressure cannot be gener-
sure in the
ated in the cuff. A leak may have occurred.
cuff
Check that cuff and pump ball are correctly
connected and that the cuff is not fitted too
loosely. Replace the batteries if necessary.
Repeat the measurement.
«ERR 5» Abnormal
The measuring signals are inaccurate and
result
no result can therefore be displayed. Read
through the checklist for performing reli-
able measurements and then repeat the
measurement.*
«HI»
Pulse or
The pressure in the cuff is too high (over
cuff pres-
300 mmHg) OR the pulse is too high (over
sure too
200 beats per minute). Relax for 5 minutes
high
and repeat the measurement.*
«LO»
Pulse too
The pulse is too low (less than 40 beats per
low
minute). Repeat the measurement.*
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.

Publicité

Table des Matières
loading

Table des Matières