OPERATION
3.3.

Displaying the instrument data

(1) By pressing the [TEST] key, the stored instrument data can be called up
( Fig. 5).
This includes:
the current gas concentration
the test gas mode
the value for the lower reaction point ("
the value for the upper reaction point ("
the minimum oxygen concentration ("
the measured peak values ("
During the display of the peak value, press the [TEST] key:
Value is deleted
CLR
"
The alarm duration in hours (hour glass, "
The IR mode
If an IR interface is detected in the test mode, the instrument bypasses
the IR mode.
If no IR interface is detected in test mode within 30 s, or the [TEST] key
is pressed, the instrument exits this mode.
16
The measured peak values can be deleted. For this, press the test
button when displaying the relevant value.
" is displayed
ALTAIR Operating Manual
LO
ALARM
" "
HI
ALARM
" "
LO
") – only for oxygen version
HI
LO
", "
")
HRS
" and numeric value)
")
")