7.3.2 Disinfection: Manual disinfection – internal
The efficacy of manual internal disinfection must be demonstrated by the manufacturer of the disinfection agent. With Faro products, use only disinfection agents that have
been released by Faro with respect to the compatibility of materials (e.g. WL-cid / made by ALPRO)
 Immediately after internal disinfection, lubricate the Faro medical device immediately with care agents Faro spray
 Follow the instructions for use of the disinfectant.

7.3.3 Disinfection: Machine disinfection - external and internal

Faro recommends thermodesinfectors in accordance with EN ISO15883-1 that are operated with alkaline cleaning agents at a pH value omax. 10
 For program settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the thermodisinfector (complying with max. pH value of 10)
 In order to prevent negative effects on the medical device, make sure that the interior and the exterior of the medical device are dry, and then lubricate immediately
with Faro Spray.
7.4 Drying
Manual Drying
Blow off the outside and inside with compressed air until water drops are no longer visible.
Automatic Drying
The drying procedure is normally part of the cleaning program of the thermodisinfector.
 Follow the instructions for use of the thermodisinfector.