Reprocessing methods according to ISO 17664
Spray the disinfectant on a cloth, then thoroughly wipe down the
medical device and leave the disinfectant to soak in according to the
instructions from the disinfectant manufacturer.
Follow the instructions for use of the disinfectant.

7.3.2 Disinfection: Manual disinfection - internal

The efficacy of manual internal disinfection must be demonstrated by the
manufacturer of the disinfection agent. With Faro products, use only dis‐
infection agents that have been released by Faro with respect to the
compatibility of materials (e.g. WL-cid / made by ALPRO).
Immediately after internal disinfection, lubricate the Faro medical
device immediately with care agents from the Faro care system.