CAUTION
Hazard from the use of handpieces equipped with electronic micromotors.
Electronic micromotors generate much more energy than conventional pneumatic turbines and motors. Given the higher torque and speed, handpieces that
are poorly serviced, damaged or used improperly can overheat which can cause serious burn injuries to the patient.
Observe the following points.
NOTE
For safety reasons, we recommend that the tool holder system be checked annually after the warranty period expires.
The following guidelines must be observed to ensure save use of the electrically driven handpieces:
▪ The service instructions for handpieces must be precisely following when using Faro spray.
▪ Before each use, the handpiece must be checked for external damage.
▪ Before each use, perform a test run with the handpiece, and watch for a typical heating and unusual noise and vibration.
▪ Immediately stop using handpieces that act unusual.
▪ Never press the pushbutton during operation. This also includes lifting the cheek or tongue!
MC01 R / RL - Instructions for use - EN