MC01 R / RL - Instructions for use - EN
7.7 Sterilisation
Sterilisation in a steam steriliser (autoclave) in accordance with EN 13060/ISO 17665-1
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
 Before each sterilisation cycle, service the medical device with Faro care products.
CAUTION
Contact corrosion due to moisture.
Damage to product.
 Immediately remove the product from the steamsteriliser after the sterilisation cycle!
The Faro medical device has a maximum temperature resistance up to138° (280.4 °F).
Depending on the available autoclave, select a suitable procedure from the following sterilisation processes:
▪ Autoclave with three times initial vacuum:
– at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4°F)
▪ Autoclave using the gravitation method:
– at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4°F) or alternatively
– at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4°F)
 Use according to the manufacturer's Instructions for Use.