Immunity test
Test / Compliance level
< 0% U
Voltage dips, short
45° steps
interruptions and
voltage variations
0% U
on power supply
70% U
input lines
< 5% U
IEC 61000-4-11
cycles
Power frequency
(50/60 Hz)
30 A/m
magnetic field
IEC 61000-4-8
Conducted RF
3 V
150 kHz to 80 Mhz
IEC 61000-4-6
6 V
in ISM & amateur radio
bands
Radiated RF
10 V/m
80 Mhz to 2,7 GHz
IEC 61000-4-3
385 MHz - 5785 MHz
test specifications
for immunity to RF
wireless communication
equipment refer to table
9 of IEC 60601-1-2:2014
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
11.4 Test specifications for immunity to RF wireless communications equipment
IEC 60601-1-2:2014 — Table 9
Test Frequency
Band
(MHz)
385
380 - 390
450
430 - 470
710
745
704 - 787
788
810
870
800 - 960
930
1551341-F
Electromagnetic environment – guidance
for 0,5 cycle at
T
Mains power quality should be that of a typical commercial or hospital
environment. If the user of this product requires continued operation during
for 1 cycles
power mains interruptions, it is recommended that the product is powered
T
from an un-interruptible power supply or a battery.
for 25 / 30 cycles
T
for 250 / 300
T
U
T
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which this product is used exceeds
the applicable RF compliance level above, this product should be observed
to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating this product.
Interference may occur in the vicinity of equipment marked with the
following symbol:
Portable and mobile RF communications equipment should be used no
closer than 30 cm to any part of this product including cables.
Service
a)
a)
(MHz)
TETRA 400
GMRS 460,
FRS 460
LTE Band 13, 17
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850, LTE
Band 5
is the a. c. mains voltage prior to application of the test level.
Maximum
Modulation
b)
power (W)
Pulse
modulation
b)
18 Hz
FM
c)
± 5 kHz
deviation
1 kHz sine
Pulse
modulation
b)
217 Hz
Pulse
modulation
b)
18 Hz
Electromagnetic compatibility (EMC)
Distance (m)
1.8
0.3
2
0.3
0.2
0.3
2
0.3
Immunity test
level (V/m)
27
28
9
28
27