5.5.1 Closing/Opening legs manually
A
1. To open the legs, pull the leg spreader handle A off and
then to the left until it locks in the notch of the bracket.
2. To close the legs, push the leg spreader handle A off
and then to the right until it locks in the notch of the
bracket.
The shifter handle MUST lock into its mounting slot
to lock the legs in the full closed position.
WARNING!
Risk of Injury
The lift could tip and endanger the patient and
assistants.
– If the leg spreader handle is NOT positioned
completely into its mounting slot, DO NOT use
the patient lift until the leg spreader handle
is properly seated and the legs of the patient
lift are locked in place.
5.6 Emergency functions
5.6.1 Performing an Emergency Stop
(Kirilos and Atlante Electric only)
1. Press the red emergency button A on the control unit
to stop the boom and patient from raising or lowering.
2. To reset, rotate the emergency button clockwise.
5.6.2 Activating an Emergency Lowering
Activating an Emergency Lowering on the Control
Unit
(Kirilos and Atlante Electric only)
If the hand control fails, the boom can be lowered by using
the circular switch for emergency release. Use a sharp
object, such as a pencil to press the button.
12
1. Lower the boom by pressing and holding button A at
the front of the control unit.
2. Stop lowering the boom by releasing the button.
Activating a Mechanical Emergency Lowering
(Kirilos and Atlante Electric only)
If the emergency lowering on the control unit is not working,
a mechanical emergency lowering is available as a back-up.
This may happen in case of partial or total power failure, or
if the battery runs down during use.
WARNING!
Risk of Injury or Damage
Repeated mechanical emergency lowering leads
to a high increase of the lowering speed.
– Only use the mechanical emergency lowering
as a back-up when the emergency lowering on
the control unit is not working.
– Do not use the mechanical emergency lowering
repeatedly in close succession.
– Always lower the patient onto a supporting
surface like a bed or a chair.
– After a mechanical emergency lowering was
required, the lift must be checked to resolve
the malfunction of the control unit.
1. Locate the emergency release knob A at the top of the
actuator piston.
2. Rotate the emergency release knob A clockwise.
The boom will lower slowly without additional
downward force on the boom.
Activating an Emergency Lowering on the
Hydraulic Pump
(Atlante Hydraulic only)
1. Turn the control valve A counterclockwise to the OPEN
position.
5.7 Charging the battery
(Kirilos and Atlante Electric only)