MEDICATION LABEL (FIGURE 2)
Medication Label Symbol Definitions:
1
Patient Name
2
Date
3
Time
4
Medication
5
Dosage
Figure 2
1
2
3
INSTRUCTIONS FOR USE
Use Aseptic Technique
FILLING THE ON-Q* PUMP: (Figure 3)
NOTE:
Follow hospital protocols and applicable regulations
for filling pump.
1. Close clamp.
2. Uncap the fill port.
3. Attach filled syringe to fill port. Invert pump as shown.
4. Grasp syringe with both hands.
5. Push down on plunger continuously until volume is
dispensed. Do not handle pump while filling, as the
syringe tip may break. Repeat as necessary. Syringe
accuracy is ±4%.
NOTE:
Filling Extension Sets are provided with larger pumps
(see product insert).
CAUTION: Do not underfill pump. Underfilling pump
may significantly increase the flow rate. Do not exceed
maximum fill volume. (Table 1)
6. Remove syringe from fill port.
7. Replace fill port cap. Label with the appropriate
pharmaceutical and patient information.
6
4
5
NOTE
: The ON-Q* Pump contains a Carry Case for
holding pump.
Figure 3
CAUTION: Do not fill less than the labeled fill volume
or exceed the maximum fill volume. (Table 1)
Table 1: Fill Volume
Labeled
Maximum Fill Vol.
Fill Vol
270 ml
400 ml
TOTAL FLOW RATE DOSING
Total Flow Rate refers to bolus + basal, which is the infusion
rate per hour. (Table 2)
Table 2: Total Flow Rate
Bolus
Refill
Dose
Time
5 ml
30 min
5 ml
60 min
Retained
Vol.
335 ml
≤ 9 ml
550 ml
≤ 15 ml
Total Flow Rate
10 ml/hr + Basal Rate
5 ml/hr + Basal Rate