WARNING: If the ONDEMAND* button does not pop
back up within 30 minutes, check position of orange
indicator:
1. If orange indicator is in the bottom position, close the
clamp. Continuous medication delivery may be occurring
which can be significantly greater than the Total Flow Rate.
or
2. If orange indicator is in the top position, something may
be impeding the flow. Check for tubing kinks, closed
clamp or patency of connected devices such as catheter or
unvented filter (verify patency) according to your standard
protocol.
ONDEMAND* DEVICE REFILL CHART
5.0
2.5
Re ll Time (minutes)
The ONDEMAND* device is available in 30 or 60 minute refill times
as labeled on the device. Refill time is approximately linear.
DURING THE INFUSION
• A change in appearance and size of the pump may not be
evident during the first 24 hours after start of infusion.
• As medication is delivered, the pump will gradually
become smaller.
• Make sure:
• Clamp is open.
• There are no kinks in the tubing.
• Filter vent is not taped or covered.
• Heat, ice or cold therapy is placed away from the flow
controller.
END OF INFUSION
• Infusion is complete when pump is no longer inflated.
• Close clamp, disconnect and dispose of pump according to
your institution's protocol.
NOTE:
If pump did not perform as expected, do not discard.
Contact I-Flow* for return product instructions at
ifloproductcomplaint@kcc.com.
8
30
60
TECHNICAL SPECIFICATIONS
DELIVERY ACCURACY: When filled to the labeled volume,
basal flow rate accuracy is ±15% and bolus dose accuracy is
+10/-20% of the labeled rates when infusion is started 0-8
hours after fill and delivering normal saline as the diluent at
22°C/72°F.
TYPICAL FLOW CURVE
The flow rate may be higher or lower at the beginning and
end of the infusion (see Figure 5).
Figure 5
Typical Flow Curve
Percent Delivery Time
NOTES:
Latex is not in fluid pathway or in contact with human.
Refer to ON-Q* Pump Latex Sensitivity Technical Bulletin at
www.iflo.com.
STORAGE CONDITIONS
Store under general warehouse conditions. Protect from light
sources and heat. Keep dry.
Nominal
100%