Tube Connection; Carrying Out A Measurement; Before The Measurement; Common Sources Of Error - Microlife BP 3AS1-2 Mode D'emploi

Masquer les pouces Voir aussi pour BP 3AS1-2:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Attention!
• After the battery warning appears, the device is blocked
until the batteries have been replaced.
• Please use «AAA» long-life or alkaline 1.5 V batteries. The
use of 1.2 V accumulators is not recommended.
• If this device is left unused for long periods, please remove
the batteries from the device.
Functional check: Press and hold the O/I button down to test all the display elements. When functioning
correctly all segments appear.

4.2. Tube connection

Cuff (1-tube cuff + T-connector):
Insert the cuff tube into the opening provided on the back
of the device, as shown in the diagram.

5. Carrying out a measurement

5.1. Before the measurement

• Avoid eating, smoking as well as all forms of exertion directly before the measurement. All these
factors influence the measurement result. Try to find time to relax by sitting in an armchair in a quiet
atmosphere for about 5 minutes before measurement.
• Remove any garment that fits closely to your upper arm.
• Always measure on the same arm (normally left).
• Attempt to carry out the measurements regularly at the same time of day, since the blood pressure
changes during the course of the day.

5.2. Common sources of error

Note: Comparable blood pressure measurements always require the same conditions!
These are normally always quiet conditions.
• All efforts by the patient to support their arm can increase blood pressure. Make sure you are in a com-
fortable, relaxed position and do not activate any of the muscles in the measurement arm during the
measurement. Use a cushion for support if necessary.
• If the wrist artery lies considerably lower (higher) than the heart, an erroneously higher (lower) blood pres-
sure will be measured! (Each 15 cm difference in height results in a measurement error of 10 mmHg!)
• Cuffs that are too narrow or too short result in false measurement values. Selecting the correct cuff is
of extraordinary importance. The cuff size is dependent upon the circumference of the arm (measured
in the centre). The permissible range is printed on the cuff. If this is not suitable for your use, please
contact your dealer. Note: Only use clinically approved Original-Cuffs!
• A loose cuff or a sideways protruding air-pocket causes false measurement values.
5

Publicité

Chapitres

Table des Matières
loading

Table des Matières