11 - TROUBLE-SHOOTING CHART
PROBLEM
POSSIBLE CAUSES SOLUTIONS
• No air or
• Flow generator off.
insufficient air
comes out of the
• Flow generator
mask.
regulated wrongly.
• Patient circuit tube
not inserted.
• The mask is noisy. • Cushion mounted
incorrectly and air
leak.
• The mask does
• Mask applied
not fit the face
incorrectly.
snugly.
• Exhaled air does
• Vent ports
not flow out and
obstructed.
difficulty inhaling.
12 - DISPOSAL INSTRUCTIONS
The components of the mask are made of the following materials and
must be disposed of in accordance with the applicable laws.
COMPONENT
Main body
Forehead support
Adjusting knob
Adjusting screw
Curved fitting
Swivel fitting
Forehead rest
Cushion
Cap
Headgear
• Switch on the flow
generator.
• Regulate the flow
generator.
• Connect the tube to
the flow generator
and mask.
• Mount the cushion
correctly.
• Reposition the
headgear.
• Remove the
obstruction.
MATERIAL
Polycarbonate
Polycarbonate
Polypropylene
Polycarbonate
Polycarbonate/Nylon
Polycarbonate
Silicone
Silicone
Silicone
Nylon/polyurethane
ENGLISH
13 - TECHNICAL DATA
MINIMUM CONTINUOUS FLOW
N.B. Due to manufacturing differences, the exhalation flow rate may
differ from the values shown (±10%).
Normative references CE marking in accordance with European
Resistance
A-weighted sound
pressure level
according to ISO 3744
A-weighted sound
power level according
to ISO 3744
Operating pressure
Temperature range
Dead space volume
inside the mask
12
Directive 93/42/EEC, class IIa and
subsequent amendments
EN ISO 14971, ISO 5356-1, ISO 10993-1,
EN ISO 17510-2
at 50 lpm: 0.1 cmH
O
2
at 100 lpm: 0.5 cmH
O
2
10,2 dB(A) ± 1,5 dB(A) Respireo N
11,5 dB(A) ± 1,5 dB(A) Respireo Primo N
18,8 dB(A) ± 1,5 dB(A) Respireo N
19,5 dB(A) ± 1,5 dB(A) Respireo Primo N
4-40 cmH
O
2
5°-40°C
114 ml