Two 6 mm (1/4") internal diameter rubber or plastic hoses from the
water supply must be fitted to the two nozzles.
These hoses must be held on the respective nozzles using hose clips to avoid
NOTE
that the tube(s) gets loose or disconnected during operation.
Cooling may be carried out either through an open circuit with
eventual discharge of the water, or using a closed-circuit cooling
system.
The water temperature must be between +15 °C and +25 °C, with an
inlet pressure between 2 and 4 bar. This allows a minimum flow of
100 l/h.
The water electrical conductance must be ≤ 500 µs/cm. When the conductance
NOTE
is higher, in closed water circuit, the use of up to 20 % of Ethyl-Glycole is
suggested.
TwisTorr 704 FS User Manual / 87-901-051-01 (A)
Technical Information
Water Cooling Kit Connection
17
307/334