Rotors And Attachments; Cleaning And Care; Disinfection; Removal Of Radioactive Contaminants - Hettich MIKRO 185 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

24.2

Rotors and Attachments

24.2.1

Cleaning and care


In order to avoid corrosion and changes in materials, the rotors and accessories have to be cleaned regularly
with soap or with a mild cleaning agent and a moist cloth. Cleaning is recommended at least once a week.
Contaminants must be removed immediately.

Ingredients of suitable detergents:
soap, anionic tensides, non-ionic tensides.

After using detergents, remove detergent residue by rinsing with water (only outside of the centrifuge) or wipe off
with a damp cloth.

The rotors and accessories must be dried directly after cleaning.

In the case of biosafety systems (for further details of available biosafety systems see chapter
"Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories"), the packing rings must be checked and
cleaned regularly (weekly).The sealing ring is to be replaced immediately upon indication of crack formation,
embrittlement or abrasive wear. In order to prevent the packing ring from twisting when opening and closing the
cover, the packing ring must be lightly rubbed with talcum powder or a rubber care product.

In order to prevent corrosion as a result of moisture between the rotor and the motor shaft, the rotor should be
disassembled and cleaned at least once a month, and the motor shaft should be lightly greased.

The rotors and accessories have to be checked monthly for wear and tear and damage due to corrosion.
Rotors and attachments may no longer be utilised upon indication of wear and tear or corrosion.

Check the firm seating of the rotor on a weekly basis.
24.2.2

Disinfection


If infectious material should get on the rotors or accessories, they must be appropriately disinfected.

Ingredients of suitable disinfectants:
ethanol, n-propanol, ethyl hexanol, anionic tensides, corrosion inhibitors.

After using disinfectants, remove disinfectant residue by rinsing with water (only outside of the centrifuge) or wipe
off with a damp cloth.

The rotors and accessories must be dried directly after disinfection.
24.2.3

Removal of radioactive contaminants


The agent must be specifically labelled as being an agent for the removal of radioactive contaminants.

Ingredients of suitable agents for removing radioactive contaminants:
anionic tensides, non-ionic tensides, polyhydrated ethanol.

After removing the radioactive contaminants, remove agent residue by rinsing with water (only outside of the
centrifuge) or wipe off with a damp cloth.

The rotors and accessories must be dried directly after removing the radioactive contaminants.
24.2.4

Rotors and accessories with limited service lives

The usage period of the rotors is limited to a certain number of run cycles (centrifugation runs). The maximum
permissible number of run cycles can be seen on the rotor.
For safety reasons, the rotor may no longer be used when the maximum allowed number of running cycles
(marked on it) has been reached.
The device is equipped with a cycle counter which counts the running cycles (centrifugation runs). For a description,
see the "Cycle counter" chapter.
AB1203DEENFRIT
Rev. 05 / 05.2020
EN
73/146

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

12031203-1

Table des Matières