Replacing the mains fuse on the microscope
Attention!
Always disconnect the mains plug!
• Switch off the microscope.
• First disconnect the power cord from the
power outlet and then disconnect it at the
microscope.
• Disconnect the power cord of the power unit
if required.
• Use a screwdriver to press on the latch (55.2)
and remove the fuse holder (55.1).
• Remove the defective fuses from the fuse
holder.
• Replace them with two new fuses of the pro-
per type.
Voltage rating
for 100 V
for 115 V
for 230 V
Note
Never use replacement fuses with a different
current rating.
• Insert the fuse holder (55.1) until the latch
clicks audibly in place.
• Subsequently connect the power cord (55.4)
to the microscope and to the mains supply.
• If required, connect the power unit to the
mains supply.
Type
2x T800 mA
2x T800 mA
2x T800 mA
The integrated transmitted-light lamp does not
work
• Make sure that the plug of the lamp cable is
firmly plugged in the corresponding socket on
the rear panel of the DM IL microscope stand.
• The halogen lamp may be defective.
Fig. 55
1 Fuse holder with voltage selector module, 2 Latch,
3 Selector module with voltage indications, 4 Mains voltage,
5 Mains connection
5
Replacing the 6 V/35 W halogen lamp
Attention!
Always disconnect the power plug!
Remove the protective sleeve of the lamp only
after inserting the lamp. Avoid finger prints,
and if there are any, always wipe them off.
Caution! Lamp and housing may be hot!
• Switch off the microscope and also the power
unit, if required.
• Disconnect the power cord of the micro-
scope and also of the power unit, if required.
2
3
4
1
59