Performance data of condensers
S 90 0.23 condenser
Suitable for lab vessels up to a height of 90 mm
and objectives with numerical apertures up to
0.50. Offers an optimal matching of light and
phase rings up to liquid levels of 35 mm in phase
contrast applications.
Applications of condensers S 90 and S 55:
Illumination
method
Bright-field
Phase
contrast
Integrated
Modulation
Contrast
S 90
Light rings/
Objectives
Accessories
25x
(with diffuser)
4x – 100x
–
5x
Phaco O
10x – 20x
Phaco 1
not
possible
S 55 0.35 condenser
Suitable for lab vessels up to a height of 55 mm
and objectives with numerical apertures up to
0.60. Offers an optimal matching of light and
phase rings up to liquid levels of 50 mm in phase
contrast applications.
S 55
Objectives Accessories
2.5x
4x – 100x
5x
10x – 20x
40x – 63x
C PLAN 10x/0.22 AP 32.2
C PLAN L 20x/0.30 D
C PLAN L 40x/0.50 D
and all objectives with
pupil position D
Light rings/
(with diffuser)
–
Phaco O
Phaco 1
Phaco 2
75