Assembling the 106* or 107* lamp housings
Note
For microscope with integrated incident-light
fluorescence facility only.
Attention!
Prior to making any assembly work, always
disconnect the power supply at the external
transformer outlet and the microscope
outlet!
Replacing or changing the halogen lamp
This procedure is only required if the lamp is not
pre-installed.
• Unscrew the screw on the lid (31.1, 32.1) with
a crosstip screwdriver and remove the lid
(Fig. 33).
• Move the collector to the front.
Note
This step is not required for the 107/2 lamp
housing.
Attention!
Leave the protective covering on the lamp
until the lamp is in its holder! Avoid making
finger prints or wipe them off immediately.
Fig. 31 Series LH 106 lamp housing
1 Screw for opening of lamp housing, 2, 3 X and Y centring of
lamp*, 4 Focusing of collector, 5, 7 Mounting screws, 6 Filter
holder (intermediate component) for filter Ø 50 mm
1
3
Fig. 32/33 Opening the LH 107 lamp housing
1 Screw for opening of lamp housing
1
1
4
1
5
2
6
7