17
Technical Information
Special Turbo-V 1K-G Navigator Controller Installation
Electrical Connection
Figure 38
The turbopump is connected to the controller through an 8-pin
connector. Pins B, C and D are the 3-phase supply to the motor, pin E
is connected to the pump ground, pins G and H are not connected,
pins A and F reserved for Agilent. If the temperature sensor is
disconnected, the pump will not start. To prevent damage to the
pump when the temperature exceeds 60 °C, the sensor automatically
cuts out the power supply.
284/294
Special Turbo-V 1K-G User Manual / 87-901-006-01 (B)