Pump Fixing - Agilent Technologies SpecialTurbo-V 1K-G Mode D'emploi

Table des Matières

Publicité

16
Instructions for Use

Pump Fixing

Pump Fixing

WARNING!
If a rotor failure occurs, the connection of the pump to the system could be
subjected to a significant torque. If the connection is not sufficient to
withstand that torque, the pump could detach from the system or the motor
housing could detach from the pump envelope. In this case metal fragments
could be projected from the pump or system, which could cause serious
injury or death and/or damage to surrounding equipment.
Fix the Special Turbo-V 1K-G in a stable position mounting the inlet
flange of the turbopump to the system counter-flange, with a
connection capable of withstanding a torque of 8900 Nm around its
axis. The following table shows, for ISO 160 F inlet flange, the
necessary number of screws and the relevant fixing torque.
Tab. 1
FLANGE
ISO 160 F
The class of the steel screws for "F" flange must be > 8.8
The Special Turbo-V 1K-G can be installed in any position.
The Special Turbo-V 1K-G cannot be fixed by means of its base.
NOTE
CAUTION!
The Special Turbo-V 1K-G belongs to the second installation (or overvoltage)
category as per directive EN 61010-1. Connect the device to a mains line that
satisfy the above category. The Special Turbo-V 1K-G has Input/Output and
serial communication connectors that must be connected to external circuits in
such a way that no electrical part is accessible. Be sure that the insulation of
the device connected to the Special Turbo-V 1K-G is adequate even in the case
of single fault as per directive EN 61010-1.
For installation of optional accessories, see "Technical Information".
234/294
Special Turbo-V 1K-G User Manual / 87-901-006-01 (B)
N.
FIXING TORQUE
8
25 Nm