Switching On And Use Of Special Turbo-V 1K-G - Agilent Technologies SpecialTurbo-V 1K-G Mode D'emploi

Table des Matières

Publicité

16
Instructions for Use
Use

Switching on and Use of Special Turbo-V 1K-G

The following instructions apply to the Special Turbo-V 1K-G used
togheter with Navigator Controller.
To switch on the Special Turbo-V 1K-G it is necessary to supply the
mains. The integrated controller automatically recognizes the
interlock and start signals presence and start up the pump.
The first pump start up is in "Soft Start" mode. When the start up
cycle is finished, the "Soft Start" mode automatically is disabled, and
the following start ups are without the "Soft Start" mode. To re-
enable the "Soft Start" mode it must be activated by the suitable
software command (see the paragraph "RS 232/485
COMMUNICATION DESCRIPTION" in the appendix "Technical
Information").
The green LED located on the Navigator controller base front panel
indicates with its flashing frequency the system operating conditions:
236/294
with no flashing: the pump is normally rotating;
slowly flashing (period of about 400 ms): the system is in ramp,
or in braking, or in Stop, or in "Waiting for interlock" status;
fast flashing (period of about 200 ms): error condition.
Special Turbo-V 1K-G User Manual / 87-901-006-01 (B)