17
Technical Information
TV 301SF Controller Installation
Connection C – Electrical
Figure 48
The turbopump is connected to the controller through an 6-pin
connector. Pins B, C and D are the 3-phase supply to the motor, pins
A and F are connected to the temperature sensor (NTC type, 30 KΩ
resistance at 25° C) and pin E is connected to the pump ground.
If the temperature sensor is disconnected, the pump will not start. To
prevent damage to the pump when the temperature exceeds 60° C,
the sensor automatically cuts out the power supply.
290/300
TV 301SF Navigator User Manual / 87-900-957-01 (H)