N.
202
203
204
205
206
300
301
302
320 to
399
400
402
500
503
504
Special Turbo-V 1K-G User Manual / 87-901-006-01 (B)
Read/
Data
Description
Write
Type
R
N
Pump power in W
(pump current x pump
voltage duty cycle)
R
N
Driving frequency in Hz
R
N
Pump temperature in °C
R
N
Pump status
R
N
Error code
R
N
Cycle time in minutes
(zeroed by the reset
command)
R
N
Cycle number (zeroed
by the reset command)
R
N
Pump life in hours
(zeroed by the reset
command)
Reserved to Agilent service
R
A
CRC EPROM (QE)
R
A
CRC Param. (PA)
Reserved to Agilent service
R/W
N
RS 485 address
R/W
L
Serial type select
Technical Information
Window-Meanings
Admitted Values
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description: see the
following figure
0 to 999999
0 to 9999
0 to 999999
QE5XXXX
(where "XXXX" are variable)
PA5XXXX
(where "XXXX" are variable)
0 to 31 (default = 0)
0 = RS 232
1 = RS 485
(default = 0)