Boston Scientific Latitude Communicateur Manuel Du Propriétaire page 52

Table des Matières

Publicité

Two Yellow Collecting Waves
Description: The Communicator started but was
Action:
48
LATITUDE
not able to complete the interrogation
within the time allowed.
• Ensure the Communicator is
optimally placed as described
in "Where to Place Your
Communicator" on page 12.
• Face the Communicator. Sit directly
in front of the Communicator. Make
sure you are within 1 meter (3 feet)
of the Communicator.
• Remain still until the interrogation is
complete. Do not move away from
the Communicator.
• Move any wireless electronic
products (such as cordless or
cellular phones or baby monitors)
that are within 1 meter (3 feet) of the
Communicator.
To verify that troubleshooting was a
success:
• Press the Heart button to start
another interrogation. If the
interrogation was a success, all
three Collecting Waves will light
green for 2 minutes.
Communicator Patient Manual

Publicité

Chapitres

Table des Matières
loading

Table des Matières