INSTRUCTIONS FOR USE
To use the autoclave, proceed in the following way:
- Open the door by turning the knob (2 fi g.1) in an anti-clockwise direction.
- Press the main luminous switch (1 fi g. 1).
USING THE MENU
The main menu consists of 3 submenus:
1 - Programs
2 - Test
3 - Options
To access the various submenus, move the selection indicator by pressing the (+) or
(-) key and then confi rming the selection with the
Press the (\\) key to cancel the operation.
SWITCHING ON FOR THE FIRST TIME
When the device is switched on for the fi rst time, the message "Add H
the display. In order to carry out the sterilization cycles, the clean water tank must be
fi lled as described below.
FILLING THE CLEAN WATER TANK
- Use only distilled water (see technical specifi cations on page 42).
- When the message "Add H
the following ways:
AUTOMATIC FILLING
- Insert the part of the pipe with the quick coupling into the water-fi ll coupling (12 fi g. 2)
and place the other end in the can.
- Press the (+) button to start the water pump.
- Press the (-) or (\\) key to stop the operation.
- Filling automatically stops once the maximum level of the tank has been reached.
MANUAL FILLING
- Insert the part of the pipe with the quick coupling into the hole (14 fi g. 3).
- Place the funnel into the other end of the pipe and pour max. 2 liters of distilled water.
INTRODUCING MATERIAL TO BE STERILIZED INTO THE CHAMBER
Package the load very carefully ensuring that the packaging chosen is compatible with the
medical devices to be sterilized.
The material used must be:
- Compatible with the sterilization process.
- Compatible with the labeling system.
- The packaging material must comply with standard EN 868.
Place the thoroughly cleaned material on the trays, distributing it uniformly.
The chamber can contain a maximum of:
2 Kg
of packaged material
4 Kg
of unpackaged material
1.3 Kg
of porous load
O" appears on the LCD, fi ll the clean water tank in one of
2
key.
O" will appear on
2
29 29